RIDASCREEN®FAST Mandel/Almond

0,01 

RIDASCREEN®FAST Almond is a sandwich enzyme immunoassay for the quantitative analysis of almond or parts of almond in food.

Category: SKU: R6901 Tags:

Description

Intended use:

RIDASCREEN®FAST Almond is a sandwich enzyme immunoassay for the quantitative analysis of almond or parts of almond in food. As representative samples for the food groups pastries and sweets the following matrices have been tested during the assay´s development: cereals, cookies, ice cream and chocolate. The assay can be likely used for analysis of other food samples too; this has to be checked by the user.

 

General Information:

The allergen can be present as an ingredient or as a contamination in raw and cooked products. According to the regulation (EU) No. 1169/2011, almond must be declared on food labels as it can induce allergic reactions. Similar regulations exist e.g. in the USA, Canada, Australia and New Zealand.

 

Benefits:

  • Widespread testing for cross reactivity (80 matrices have been evaluated).
  • Accuracy: recovery of spiked samples was between 66 – 124 % (mean: 97.8 %).
  • Validated for automates.

 

Specifications
Article Numbers R6901
Test format Microtiter plate with 48 wells (6 strips with 8 removable wells each)
Sample preparation Homogenization, extraction and centrifugation
Incubation time 30 min
Room temperature
LOD (Detection limit) 0.1 mg / kg (ppm) almond (mean)
0.0 – 0.23 mg / kg (ppm) almond*
*depending on matrix
LOQ (Limit of quantification) 2.5 – 20 mg / kg (ppm) almond
(depending on matrix, level of food processing and extraction method)
Cross reactivity Cross reaction to apricot stone, mulberries, mahaleb cherry (cross reactions to other plants from genus Prunus are possible).
Validated matrices Representative samples for the food groups pastries and sweets: cereals, cookies, ice cream and chocolate.
Available application notes
  • More sensitive analysis (improved LOQ of 0.625 mg / kg (ppm)).
Detected analyte Proteins from almond
Evaluation Microtiter plate spectrophotometer (450 nm)